Model 40.10 special features:

  • Automatic temperature calibration, (effected automatically before each distillation run without the need of external calibration equipment).
  • Level follower with stepper motor and
  • Automatic calibration of the volume measurement, zero point and/or 100 ml.
  • Low mass / low volt heating.
  • Accuracy: +/- 0,1 ml volume, +/- 0,01°C temperature.
  • Level follower with cooling/heating system, range 0 – 60°C.
  • Distillation range 0 – 450°C.
  • Calculation of cetane index.
  • 100and 200 ml measuring cylinders useable, consequently 100 and 200 ml test samples may be distilled.
  • "door open" sensor.
  • built-in environment-friendly Peltier elements for cooling and heating.
  • built-in automatic fire-extinguishing system for connection to an external Nitrogen (N2) supply.
  • automatic-dry-point-detection-system for solvents, including thermocouple, 200 ml flask, PTFE plug, and accessories.
  • Emergency ”STOP ALL” switch.
  • New, slim design.
  • Learning Program.
  • Built-in, Windows-based PC, with 10,4“ TFT active color LCD-Display with Touch-Screen for a most comfortable operation of the analyzer, including:

              – Intel Atom microprocessor.

              – 16 GB compact flash card.

              – Ethernet interface for intranet connection.

              – USB plugs for possible connections of:

              – Keyboard.

              – Mouse / trackball.

              – Printer.

              – Software package.

              – Windows software.

NEW Automatic Distillation Apparatus Model 40.10

   

with installed microprocessor control and integrated MS-Windows 7 PC used for distillation products, solvents, and crude oil in accordance the following standards:

ASTM D86 (group 0-4), ASTM D216, ASTM D285 (modified), ASTM D447, ASTM D1078 (internal standardized with solvents and group 7 and 8), IP 123, ISO 3405 and DIN 51751.

Multiple units may be linked to a central PC.

The units are programmed with all necessary distillation parameters. All units can operate independently.

All relevant distillation data is stored on the built-in PC and can optionally be stored on a network device, e.g. a server, NAS or dedicated storage device or USB devices.

Example of different distillations:

ASTM D86 group 0,1,2,3: Distillation of hydrocarbon up to 300°C FBP, e.g. all type of petrol, aviation, turbine fuel etc.
ASTM D86 group 4: Distillation of diesel fuel, gas oil, furnace oil, heavy product etc, up to 450°C FBP.
ASTM D1078 group solvents. (visual - manual dry point-detection).
ASTM D1078 group solvents (automatic dry point detection).
ASTM D1078 group 7 and 8 (modified dry point detection).
ASTM D285 (modified), group 6: Distillation of crude oil

 

Main voltage: 110 ~ 235 V AC, 50 ~ 60 Hz, ± 10% (wide range input).
Operation voltage: 24 V DC.
Connecting cable: approx. 2 m long.
Required power input: approx. 2,0 kW.
Recommended main fuse: 16 A lag.
Dimensions approx.: width 515 mm, depth 500 mm, height 670 mm.
Weight: approx. 56,0 kg.
Ambient Temperature: 16°C ~ 32°C (60°F ~ 90°F).
Humidity: 10 % – 65%.

 

Each of our analyzers will be delivered with the following standard accessories

for immediately operation:

 – Distillation flask 125 ml

–  Distillation flask 200 ml

–  Ceran plate with hole 32 mm [1]

–  Ceran plate with hole 37,5 mm [1]

–  Ceran plate with hole 50 mm [1]

–  Resistance thermometer Pt 100 with cable and plug.

–  Special graduated receiver cylinder complete with metal base.

–  Wiper for condenser tube.